10 eum, 11 fil, 12 fsr – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 37

Advertising
background image

tinb07e1-e 2011-07 Vacuum Gauge Controller

37

bined by the logical operator OR. Example: 1001 =
Device error and syntax error.

6.3.10 EUM

Switch the emmision. See Chapter 5.2.10 Emission
(EMi),  26
.

S: EUM[,a,b,c]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,b,c<CR><LF>

6.3.11 FIL

Filter. Measurement filter. See Chapter 5.2.1 Measure-
ment filter (FiLt),  23
.

S: FIL[,a,b,c]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,b,c<CR><LF>

6.3.12 FSR

Full scale range. Full scale range of linear sensors
(CDG). Se
e Chapter 5.2.3 Measuring range (FS),  24.

S: FSR[,a,b,c]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,b,c<CR><LF>

Parameter

Significance

a

Emission for channel 1

0 = Manually
1 = Automatic (default)

b

Emission for channel 2
(see above)

c

Emission for channel 3
(see above)

Parameter

Significance

a

Filter for channel 1

0 = Fast
1 = Medium (default)
2 = Slow

b

Filter for channel 2
(see above)

c

Filter for channel 3
(see above)

Parameter

Significance

a

Full scale range of sensor 1

0 = 0.01 mbar
1 = 0.01 Torr
2= 0.02 Torr
3 = 0.05 Torr

4 = 0.10 mbar
5 = 0.10 Torr
6 = 0.25 mbar
7 = 0.25 Torr
8 = 0.50 mbar
9 = 0.50 Torr

10 = 1 mbar
11 = 1 Torr
12 = 2 mbar
13 = 2 Torr
14 = 5 mbar
15 = 5 Torr

16 = 10 mbar
17 = 10 Torr
18 = 20 mbar
19 = 20 Torr
20 = 50 mbar
21 = 50 Torr

22 = 100 mbar
23 = 100 Torr
24 = 200 mbar
25 = 200 Torr
26 = 500 mbar
27 = 500 Torr

28 = 1000 mbar
29 = 1100 mbar
30 = 1000 Torr

31 = 2 bar
32 = 5 bar
33 = 10 bar
34 = 50 bar

b

Full scale range of sensor 2
(see above)

c

Full scale range of sensor 3
(see above)

Advertising