1 switching function relay, 2 error signal relay, 5 outputs – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 9: 1 analog output, 2 recorder output, 3 computer interface, 6 scope of delivery, Switching function relay, Error signal relay, Outputs

Advertising
background image

tinb07e1-e 2011-07 Vacuum Gauge Controller

9

2.4.1

Switching function relay

2.4.2

Error signal relay

2.5 Outputs

2.5.1

Analog output

2.5.2

Recorder output

2.5.3

Computer interface

2.6 Scope of delivery

Contact type

Change-over contact, float-
ing

Load (ohmic)

Max. 60 VDC, 0.5 A
Max. 30 VAC, 1 A

Lifetime

Mechanical:
10

7

cycles

Electrical:
10

5

cycles at maximum load

Connection

D-sub, 25 pins, female.
See Fig. 3-8,  13.

Number

1

Delay time

≤ 20 ms

Contact type

Change-over contact, float-
ing

Load (ohmic)

Max. 60 VDC, 0.5 A
Max. 30 VAC, 1 A

Lifetime

Mechanical:
10

7

cycles

Electrical:
10

5

cycles at maximum load

Connection

D-sub, 25 pins, female.
See Fig. 3-8,  13.

Number

1 per channel

Voltage range

0…10 VDC

Deviation from
displayed value

± 50 mV

Internal resistance

47

Ω

Relation between
voltage and pressure

Sensor dependent

Connection

D-Sub, 9 pins, male.
See Fig. 3-9,  14.

Number

1

Voltage range

0…10 VDC

Resolution

1 mV

Accuracy

± 20 mV

Internal resistance

3300

Ω

Relation between
voltage and pressure

Programmable

Connection

D-Sub, 9 pins, male.
See Fig. 3-9,  14.

Default

RS232C

Protocol

ACK/NAK

ASCII with 3-character mne-
monics. Bidirectional data
flow.

Signals

Only TXD and RXD used

Baud rate

9600, 19200, 38400

Connection

D-Sub, 9 pins, female.
See Fig. 3-10,  14.

Designation

Number

Vacuum Gauge Control-
ler

1

Mains cable

1

Rubber strip
Rubber feet

1
2

Collar screws
Plastic sleeves

4
4

CD-ROM manual

1

EC Declaration of Confor-
mity

1

Installation manual

1

Advertising
This manual is related to the following products: