36 tlc, 37 tra, 38 trs – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 44: 39 uni, 40 wdt

Advertising
background image

44

tinb07e1-e 2011-07 Vacuum Gauge Controller

6.3.36 TLC

Torr lock. See Chapter 5.4.4 Torr lock (tr-L),  30.

S: TLC[,a]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a<CR><LF>

6.3.37 TRA

Test RAM. Test the RAM. See Chapter 5.4.6 RAM test
(rA-t),  30
.

S: TRA<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: aaaa<CR><LF>

The control character <ENQ> starts the test. It takes
approximately one second to complete the test.

6.3.38 TRS

Test RS232C interface. Test the RS232C interface. See
Chapter 5.4.13 RS232C test (rS-t),  31.

S: TRS<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

The control character <ENQ> starts the test. The test can
be stopped by pressing Ctrl-C.

6.3.39 UNI

Unit of measurement. See Chapter 5.3.1 Unit of mea-
surement (unit),  27
.

S: UNI[,a]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a<CR><LF>

6.3.40 WDT

Watchdog control. See Chapter 5.4.3 Watchdog control
(dt-C),  29.

S: WDT[,a]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a<CR><LF>

Parameter

Significance

a

Torr lock

0 = Off (default)
1 = On

Parameter

Significance

aaaa

Error status. See Chapter 6.3.9
ERR,  36
.

Parameter

Significance

a

Unit of measurement

0 = mbar/bar (default)
1 = Torr
2 = Pascal
3 = Micron

Parameter

Significance

a

Error acknowledgement

0 = Manually
1 = Automatic (default)

Advertising