Tab. 4-2,  18 – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 18

Advertising
background image

18

tinb07e1-e 2011-07 Vacuum Gauge Controller

4.4.3.4

Switching degas function on

The degas function of the following sensors can be
switched on manually: BPG, BCG.

1

Press the CH button to select the required channel

2

Keep the UP button pressed for approximately 1
second. See Fig. 4-4,  17.

The degas function of the sensor on the selected channel
is switched on. The DEG indicator is illuminated.

4.4.3.5

Switching high-vacuum circuit off

The high-vacuum circuit of the following sensors can be
switched off manually: PEG.

For this purpose the sensor control must be set to HAnd.
See Chapter 5.2.8 Sensor deactivation (S-oFF),  25.

1

Press the CH button to select the required channel

2

Keep the DOWN button pressed for approximately
1 second

Fig. 4-5

Press the DOWN button for 1 second

The sensor on the selected channel is switched off. The
HV indicator is dark. The display shows the status oFF.

4.4.3.6

Switching degas function off

The degas function of the following sensors can be
switched off manually: BPG, BCG.

1

Press the CH button to select the required channel

2

Keep the DOWN button pressed for approximately
1 second. See Fig. 4-5,  18.

The degas function of the sensor on the selected channel
is switched off. The DEG indicator is dark.

4.4.3.7

Identifying a sensor

1

Keep the UP and DOWN buttons pressed for
approximately 1 second

Fig. 4-6

Press the UP and DOWN buttons for 1 second

The display area of the individual sensor shows the con-
nected sensors. See Tab. 4-2,  18.

Tab. 4-2

Sensor identification

NOTE:
In the case of BPG/BCG/HPG sensors, the software
version number of the sensor is also shown (e.g.
1.20).

para

ch

Display

Significance

PSG

Pirani gauge (PSG)

PCG

Combined gauge
Pirani / Capacitance (PCG)

PEG

Cold cathode gauge (PEG)

MPG

Combined gauge
Cold cathode / Pirani (MPG)

CDG

Capacitive analog gauge, linear
(CDG025, CDG045, CDG100)

CDG d

Capacitive digital gauge, linear
(CDG025D, CDG045D, CDG100D,
CDG160D)

bPG

Combined gauge
Hot ionization / Pirani (BPG400)

bPG2

Combined gauge
Hot ionization / Pirani (BPG402)

bCG

Combined gauge
Hot ionization / Pirani / Capacitance

HPG

Combined gauge
Hot ionization / Pirani (HPG)

noSEn

No sensor found

noid

No sensor identification found

para

ch

Advertising