S: LOC[,a]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a<CR><LF>

6.3.18 OFC

Offset correction. Offset correction function for linear sen-
sors (CDG).
See Chapter 5.2.4 Offset (oFS),  24.

S: OFC[,a,b,c]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,b,c<CR><LF>

6.3.19 OFD

Offset display. Offset correction values for linear sensors
(CDG).
See Chapter 5.2.4 Offset (oFS),  24.

S: OFD[,±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc]
<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: ±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc<CR><LF>

6.3.20 PNR

Program number. Firmware version number. See
Chapter 5.4.2 Firmware version (Pnr),  29.

S: PNR<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a<CR><LF>

6.3.21 PR1

Pressure sensor 1. Pressure reading of sensor 1.

S: PR1<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,±b.bbbbE±bb<CR><LF>

NOTE:
The commands PR2 and PR3 concern the sensors 2
and 3, respectively. The commands are analogous to
the PR1 command.

Parameter

Significance

a

Parameter setup lock

0 = Off (default)
1 = On

Parameter

Significance

a

Offset correction of channel 1

0 = Off (default)
1 = On
2 = Determine the offset value and
activate offset correction function
3 = Adjust the zero point of a
CDGxxxD

b

Offset correction of channel 2
(see above)

c

Offset correction of channel 3
(see above)

Parameter

Significance

±a.aaaaE±aa

Offset value of sensor 1 in current
unit of measurement

Default: 0.0000E+00

±b.bbbbE±bb

Offset value of sensor 2 in current
unit of measurement (see above)

±c.ccccE±cc

Offset value of sensor 3 in current
unit of measurement (see above)

Parameter

Significance

a

Firmware version

Example: 302-534-D

Parameter

Significance

a

Status of channel 1

0 = Measurement data ok
1 = Underrange
2 = Overrange
3 = Sensor error
4 = Sensor switched off
5 = No sensor
6 = Identification error
7 = BPG/HPG error

±b.bbbbE±bb

Pressure reading of sensor 1 in the
current unit of measurement

Parameter

Significance