tinb07e1-e 2011-07 Vacuum Gauge Controller

43

6.3.33 TID

Transmitter identification. Sensor identification. See
Chapter 4.4.3.7 Identifying a sensor,  18.

S: TID<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,b,c<CR><LF>

6.3.34 TIO

Test I/O. Test the relays. This command allows to switch
a single relay or several relays at a time.

S: TIO[,a,bb]<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: a,bb<CR><LF>

NOTE:
The relay status is a hexadecimal number. It may be
combined by the logical operator OR. Example: 24 =
Switching functions relays 3 and 6 on.

6.3.35 TKB

Test keyboard. Test the keyboard.

S: TKB<CR>[<LF>]

R: <ACK><CR><LF>

S: <ENQ>

R: aaaa<CR><LF>

The control character <ENQ> starts the test. The Vac-
uum Gauge Controller polls the keyboard and sends a
message to the computer.

NOTE:
The keyboard status is a binary number. It may be
combined by the logical operator OR. Example: 0011
= DOWN and UP pressed at the same time.

Parameter

Significance

a

Identification of sensor 1

PSG
PCG
PEG
MPG
CDG
BPG
BPG402
BCG
HPG
noSen
noid

b

Identification of sensor 2
(see above)

c

Identification of sensor 3
(see above)

CAUTION

Relay test.

In this test the relays switch irrespective of the
actual pressure. This may cause unintended
switching of devices.

Unplug the RELAY connection before per-
forming a relay test.

Parameter

Significance

a

Test status

0 = Off
1 = On

bb

Relay status

00 = All relays off
01 = Switching function 1 relay on
02 = Switching function 2 relay on
04 = Switching function 3 relay on
08 = Switching function 4 relay on
10 = Switching function 5 relay on
20 = Switching function 6 relay on
40 = Error signal relay on
7F = All relays on

Parameter

Significance

aaaa

Keyboard status

0000 = No button pressed
1000 = CH pressed
0100 = PARA pressed
0010 = DOWN pressed
0001 = UP pressed