THE "NET" LED indicates the network status of the CDGxxxD gauge:

LED

Status

Off

Device is off

Red

Error. The alarm bit in Exception Status is
set. See section 7.1.11.

Green/flashing

Device is ok and uses acyclic data traffic.
See section 5.1.

Green

Device is ok and uses cyclic data traffic.
See section 5.2.

The gauge supports all data rates defined in the IEC 61158 Type 3 / IEC 61784
standards (

→ [5]) up to 12 Mbaud. Automatic data rate setting is supported.

Alternatively, a fixed data rate can be selected.

The ident number assigned to the gauge by the PNO (

→ [4]) is:

Gauge

Ident number (hexadecimal)

CDGxxxD 0B7E


Depending on the standard telegrams used (

→ section "Cyclic Message Tele-

grams"), the following configuration data have to be transmitted to the gauge during
the configuration phase:

Standard

telegram

M

S

Standard

telegram

S

M

M

S

S

M

Configuration data

ParamChannel

ParamChannel

Exception Status

Process Value UINT

16

Process Value float

Channel Status

Trip Relay 1-2

1

x

x

0x42,0x82,0x05,0x03

2

x x

0x42,0x84,0x05,0x08

2

7

x x x x 0xC4,0x87,0x8A,0x0A,0x0A,0x05,0x03

2

8

x x x x

0xC4,0x87,0x8C,0x0A,0x0A,0x05,0x08

201

x x x x

0x44,0x85,0x05,0x03,0x05,0x0A

202

x x x x

0x44,0x87,0x05,0x08,0x05,0x0A

2

203

x x x x x x

0xC6,0x87,0x8D,0x0A,0x0A,0x05,0x03,0x05,0x0A

2

204

x x x x x x

0xC6,0x87,0x8D,0x0A,0x0A,0x05,0x08,0x05,0x0A


3.4 "NET" LED

3.5 Data Rate

3.6 Ident Number

3.7 Configuration Data