INFICON SKY CDGxxxD (Profibus) User Manual
Page 43
tira54e1-a (2009-01) CDGxxxD.cp
43
This parameter contains the following information:
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Byte
0
0 0 0 0 0 0 0 NT
Byte
1
0 0 0 0 0 0
EW
0
• NT = Not At Temperature
This bit is set to "1" as long as the specified nominal sensor temperature is not
reached.
• EW = Electronics Warning
This bit is set to "1" if a Temperature Sensor error occurs.
This parameter contains the nominal temperature in degrees centrigrade (Celsius)
to which the sensor warms up.
This parameter defines a customer specified offset which will be added to the
Offset-A value from the Zero-Adjust Service in terms of the currently selected data
type (ID 21) and data unit (ID 22).
This parameter defines the filter function of the basic device.
Filter
Description
0 Dynamic
1 Fast
2 Slow
3…255 Reserved
12.1.4 Sensor Warning (ID 104)
12.1.5 Sensor
Temperature
(ID 105)
12.1.6 Offset (ID 201)
12.1.7 Filter (ID 202)