If one of the bits 0…2 is set in the Exception Status, the current error can be read
in the "Exception Detail Alarm" parameter. The "Exception Detail Alarm" parameter
consists of a total of 8 bytes that inform on the error status of the gauge.

Due to the use of the extended error output format, these bytes have the following
structure:

Byte

no

Name

Description

Value

Common Exception Detail Alarm

0

Size

Number of subsequent bytes
used for description of the alarm

2

1 Detail

0

Error message
(simple, UINT8, 1 byte)

See table below

2 Detail

1

Error message
(simple, UINT8, 1 byte)

See table below

Device Exception Detail Alarm

3

Size

Number of subsequent bytes
used for description of the alarm

2

4
5

Detail Error

message

(array, UINT8, 2 bytes)

Sensor Alarm

(ID 103) of CDG

Transducer Block

(

→ section 12.1.3)

Manufacturer Exception Detail Alarm

6

Size

Number of subsequent bytes
used for description of the alarm

1

7 Detail

Error message
(simple, UINT8, 1 byte)

See table below

Bit

Common Exception Detail 0

Bit

Common Exception Detail 1

0

Internal diagnostic exception

0

0

1 Microprocessor

exception

1

0

2 EPROM

exception

2

0

3 EEPROM

exception

3

0

4 RAM

exception

4

0

5 Communications

exception

5

0

6

Internal real-time exception

6

0

7 0

7 0

Bit Manufacturer Exception Detail

0 Internal

communication

exception

1 Incompatible

software

2 0

3 0

4 0

5 0

6 0

7 0


7.1.12 Exception Detail Alarm