INFICON SKY CDGxxxD (Profibus) User Manual

Page 23

Advertising
background image

tira54e1-a (2009-01) CDGxxxD.cp

23

If one of bits 4…6 is set in the Exception Status, the current warning can be read in
the parameter "Exception Detail Warning". The Exception Detail Warning pa-
rameter consists of a total of 9 bytes that inform on the error status of the gauge.

Due to the use of the extended error output format, these bytes have the following
structure:

Byte

no

Name

Description

Value

Common Exception Detail Warning

0

Size

Number of subsequent bytes
used for description of the
warning

2

1 Detail

0

Error message
(simple, UINT8, 1 byte)

See table above

2 Detail

1

Error message
(simple, UINT8, 1 byte)

See table above

Device Exception Detail Warning

3

Size

Number of subsequent bytes
used for description of the
warning

3

4 Status

extension Status extension

(simple, UINT8, 1 byte)

Status extension

(ID 102) of CDG

Transducer Block

5
6

Detail Error

message

(array, UINT8, 2 bytes)

Sensor Warning

(ID 104) of CDG

Transducer Block

Manufacturer Exception Detail Warning

7

Size

Number of subsequent bytes
used for description of the
warning

1

8 Detail

Error message
(simple, UINT8, 1 byte)

See table above


This parameter identifies the number of hours that the basic device has been
powered ON. The parameter has a resolution of 1 hour.

This proprietary parameter describes the firmware version of the basic device.

This proprietary parameter describes the version of the basic device hardware.

This proprietary parameter allows access of the corresponding part (Detail 0 and
Detail 1) of the Common Exception Detail Alarm parameter (ID 27). See section
7.1.12.

This proprietary parameter allows access of the corresponding part (Detail) of the
Device Exception Detail Alarm parameter (ID 27). See section 7.1.12.

This proprietary parameter allows access of the corresponding part (Detail) of the
Manufacturer Exception Detail Alarm parameter (ID 27). See section 7.1.12.

This proprietary parameter allows access of the corresponding part (Detail 0 and
Detail 1) of the Common Exception Detail Warning parameter (ID 28). See section
7.1.13.

7.1.13 Exception

Detail

Warning (ID 28)

7.1.14 Run hours (ID 36)

7.1.15 Basic Device Firmware

Revision Level (ID 202)

7.1.16 Basic Device Hardware

Revision Level (ID 203)

7.1.17 Common

Exception

Detail Alarm (ID 204)

7.1.18 Device Exception Detail

Alarm (ID 205)

7.1.19 Manufacturer

Exception

Detail Alarm (ID 206)

7.1.20 Common

Exception

Detail Warning (ID 207)