INFICON SKY CDGxxxD (Profibus) User Manual

Page 45

Advertising
background image

tira54e1-a (2009-01) CDGxxxD.cp

45


ID

Name

Structure

Data Type

Bytes

Access

Store

14 Atmosphere

Sensor Block

Command

Array UINT8 2

1_R/W
2_R/W

N

Byte

Name

Structure

Data Type

Bytes

Access

Store

0 Transition

Command

Simple UINT8 1 1_R/W

2_R/W

N

1 Target

Value Simple

UINT8

1 1_R/W

2_R/W

N

Transition Command

Name

Description

0…127

Reserved

Reserved, no action

128 User

ATM

adjust

"User Atmosphere Adjust" command

129…255

Reserved

Reserved, no action

There are no state transitions associated with the invocation of this service. It is
therefore mandatory for the user to switch the device into the desired configuration
prior to and during execution of this service. This generally means exposing the
sensor to a known environment and treating the values read during execution of
the services accordingly.

A successful service response indicates that the service was accepted and the
application process started.

To perform the "User ATM Adjust Request", vent the gauge to atmosphere and
then start the "User Atmosphere Adjust Service" without any target value.

12.2.5 Analog Input Block

Adjust Command
(ID 14)