2 device block, device behavior, Cyclic data traffic acyclic data traffic – INFICON HPG400-SP ATM to High-Vacuum Gauge (Profibus) User Manual

Page 22

Advertising
background image

22

tira36e1-a (0310) BPG/HPG400 v1.cp

The BPG400-SP / HPG400-SP behaves as described in the status diagram below.

NORMAL

RUNNING

OK

SELFTESTING

EXECUTING

SELFTEST

EXCEPTION

ABORT

(0)

CRITICAL

FAULT

INIT

IDLE

(1

4)

(1

3)

(1

0)

(8

)

(5)

(9)

(4)

(1

3)

(13)

(2

)

(1)

(3

)

After the start, the gauge independently runs through the INIT and SELFTESTING
status and eventually changes to the IDLE status (if there is no error) or to the
SELFTEST_EXCEPTION status (if there is a gauge error).

When data traffic is taken up, a difference has to be made between cyclic and
acyclic data traffic.

As soon as cyclic data interchange is taken up, the gauge automatically changes to
the EXECUTING status.

In acyclic data traffic, a START service has to be transmitted to bring the gauge to
the EXECUTING status.

3.1.2 Device Block, Device

Behavior

Cyclic data traffic

Acyclic data traffic