INFICON HPG400-SP ATM to High-Vacuum Gauge (Profibus) User Manual

Page 28

Advertising
background image

28

tira36e1-a (0310) BPG/HPG400 v1.cp

Pressure unit

Underrange

COUNTS

27 000

Torr

7.50053×10

-3

mTorr (Micron)

7.50053

mbar

1.0×10

-2

Pascal

1

Instance 2 of the Analog Sensor Input Function Block describes the functionality of
the BA/HP measuring part of the gauge.
In Instance 2, the same attributes as in Instance 1 are supported:

ID

Name

Structure Data type

Bytes Access

Store

19

Process Value
(PV)

Simple

According to
Parameter Data
Type

-

0_XI
1_R
2_R

D

20

Status

Simple

Unsigned8

1

0_XI
1/2_R

D

21

Data Type

Simple

Unsigned8

1

2_R/W

N

22

Data Units

Simple

Unsigned16

2

2_R/W

N

23

Reading Valid

Simple

Boolean

1

1_R
2_R

D

24

Full Scale

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

39

Safe State

Simple

Unsigned8

1

1/2_R/W

N

40

Safe Value

Simple

According to
Data Type value
(parameter 21)

-

1/2_R/W

N

44

Overrange

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

45

Underrange

Simple

According to
Data Type value
(parameter 21)

-

1/2_R

N

The Process Value contains the measured value of the BA/HP Device Instance in
the currently selected data unit (ID 22) and in the selected data type (ID 21).
If the device is not in the EXECUTING status (ID 25, Device Block), the value
defined in the Safe State is output.

Values output in the data unit COUNTS can be converted into a pressure value by
means of the following formulas:

HGP400-SP

3.2.3 Analog Sensor Input

Function Block
Instance 2 / SLOT 1

3.2.3.1 Process Value

Advertising