3 transducer block – INFICON HPG400-SP ATM to High-Vacuum Gauge (Profibus) User Manual

Page 33

Advertising
background image

tira36e1-a (0310) BPG/HPG400 v1.cp

33

If the pressure drops below the set threshold, the relay is activated (normally open
contact closed). If the pressure than rises above the set threshold with a hysteresis
of 10%, the relay is deactivated again (normally open contact open).

Bit

Definition

0

0

1

Low Alarm Exception: 0 = cleared; 1 = set

2

0

3

Low Warning Exception: 0 = cleared; 1 = set

4

0

5

0

6

0

7

0

Description

→ Instance 1.

Description

→ Instance 1.

This parameter is always on "1".

ID

Name

Structure Data type

Bytes Access Store

120

One Of N Status
Extension

Simple

UINT8

1

1_R
2_R

V

This parameter indicates whether the overrange or underrange of the gauge is
exceeded.

Bit7

Bit6

Bit5

Bit4

Bit3

Bit2

Bit1

Bit0

0

0

0

0

0

Underrange
Exceeded

Overrange
Exceeded

Reading
Invalid

If the gauge is operated in its overrange or underrange, the corresponding bit and
additionally the bit "Reading Invalid" is set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.

ID

Name

Structure Data type

Bytes Access Store

101

Block Type

Simple

Octet string

4

2_R

102

Status Extension

Simple

UINT8

1

1/2_R

V

103

Sensor Alarm

Struct

Array of 2 bytes

2

1/2_R

V

104

Sensor Warning

Struct

Array of 2 bytes

2

1/2_R

V

140

Full Scale State

Simple

UINT8

1

1/2_R

V

According to the table in Appendix A, the Block Type ID has the value "13".

3.2.4.2 Status

3.2.4.3 Data Type

3.2.4.4 Data Unit

3.2.4.5 Reading Valid

3.3 Transducer Block

3.3.1 One Of N Vacuum Gauge

Transducer Block /
SLOT 1

3.3.1.1 One Of N Status Extension

3.3.2 Heat Transfer Vacuum

Gauge Transducer
Block / SLOT 1

3.3.2.1 Block Type

Advertising