Calculation of the pressure (PV is the abbreviation used for Process Value):

PV

mbar

=

10

(COUNTS / 2000) - 12.5

PV

Torr

=

0.75006168 × PV

mbar

PV

Micron

=

10

-3

× PV

Torr

PV

Pa

=

100 × PV

mbar

Calculation of the pressure (PV is the abbreviation used for Process Value):

PV

mbar

=

10

(6 × COUNTS / 16000) - 9.125

PV

Torr

=

0.75006168 × PV

mbar

PV

Micron

=

10

-3

× PV

Torr

PV

Pa

=

100 × PV

mbar

This parameter is remains on "0".

Two data types are supported: Float and Integer16.
In cyclic data traffic, the data type cannot be modified. The data type setting can
only be modified when the gauge is in the IDLE status. By defining the configura-
tion data for cyclic data traffic (selection of standard telegram), the data type used
in the selected standard telegram is taken over. All settings previously made in
acyclic data traffic are thus overwritten.

If the data type is set in one instance, that data type setting applies to all instances.
Likewise, when a standard telegram is selected, the data type used by that
standard telegram will be valid for all instances

Coding

Data type

3

Integer16

8

Float

The gauge supports the following pressure units:

Coding

Pressure unit

1001

COUNTS

1301

Torr

1302

mTorr (Micron)

1308

mbar

1309

Pascal

For safety reasons, it is not possible to change the pressure unit while
the gauge is cyclically interchanging data with a DP/V0 master.
The data unit setting can only be modified when the gauge is in the IDLE
status.
In cyclic data traffic, the data unit must be set in the User Parameter
Data. All settings previously made in acyclic data traffic are overwritten
(

→ section "User Parameter Data").

If the data unit is set in one instance, that data unit setting applies to all
instances. Likewise, the data unit setting made in the User Parameter
Data is valid for all instances.

BPG400-SP

HPG400-SP

3.2.3.2 Status

3.2.3.3 Data Type

3.2.3.4 Data Unit