Instances 3 and 4 of the Analog Sensor Input Function Block describe the func-
tionality of the two switching functions (Setpoint A / Setpoint B) of the gauge
(setting the switching functions

→ [2], [3] for BPG400-SP and [5], [6] for

HPG400-SP):

Instance

Setpoint

3

A

4

B

With Process Values (ID 19), the current threshold setting (made by means of the
potentiometers) is read; with Status, the relay status (open/closed) can be read.

ID

Name

Structure Data type

Bytes

Access

Store

19

Process Value
(PV)

Simple

Unsigned16

-

1_R
2_R

D

20

Status

Simple

Unsigned8

1

1_R
2_R

D

21

Data Type

Simple

Unsigned8

1

2_R
2_W

N

22

Data Units

Simple

Unsigned16

2

2_R
2_W

N

23

Reading Valid

Simple

Boolean

1

1_R
2_R

D

The Process Value contains the current setting of the threshold potentiometers for
"Setpoint A" and "Setpoint B" in the currently selected data unit and data type.
If the pressure drops below the set threshold, the relay is closed. If after that, the
pressure rises above that threshold with a hysteresis of 10%, the relay is opened
again (

→ [2], [3] for BPG400-SP and [5], [6] for HPG400-SP).

The Process Value PV is output in the currently selected data unit and data type.

Values output in the pressure unit COUNTS can be converted into a pressure value
by means of the following formulas:

p

mbar

=

10

(COUNTS / 2000) - 12.5

p

mbar

=

10

(COUNTS - 22 999) / 2444

The threshold voltages of the Setpoint potentiometer converted into a pressure
value by means of the following formulas:

p

mbar

=

10

(1.23011 × U) - 9.30102999

p

mbar

=

10

(9 × U / 10) - 6

Where:

U is the threshold voltage [V] of the corresponding Setpoint (A, B),
measured at the D-Sub connector (sensor cable connector) of the gauge
(

→ [2], [3] for BPG400-SP and [5], [6] for HPG400-SP).

The switching functions work only at pressures <100 mbar. If a higher
threshold has been selected, the relay is only activated at pressures
lower than 100 mbar.

3.2.4 Analog Sensor Input

Function Block
Instances 3
and 4 / SLOT 1

3.2.4.1 Process Value

BPG400-SP

HPG400-SP

BPG400-SP

HPG400-SP