This parameter indicates the emission status of the gauge.

Emission status

Bit 1

Bit 0 Meaning

False

0

0

Emission Off

TRUE

0

1

Emission current low

TRUE

1

0

Emission current high

TRUE

1

1

Degas

Emission status

Bit 0 Meaning

FALSE

0

Emission Off

TRUE

1

Emission On

This parameter indicates the value of the emission current in mA.

Emission current [mA]

Bit 1

Bit 0

0

0

0

0.025

0

1

5

1

0

20

1

1

This parameter indicates whether the gauge is in Degas mode.

Degas Status

Bit 0

Bit 1 Meaning

FALSE

0

0

Degas Off

TRUE

1

1

Degas On

This service is used for activating the Degas mode via Profibus.

Byte Name

Structure Data type

Bytes

Access

0

State Command

Simple

Unsigned8

1

1/2_W

1

State Command
Data Field...

1)

Simple

Unsigned8

1

1/2_W

State

Command

Name

Description

0

Inactive

No action

1

Set Degas
State

Activates/deactivates the Degas mode

This service is used for activating/deactivating the
Degas mode (p < 7.2

×10

-6

mbar).

If the Degas mode has not been deactivated with
Degas Off before, it is automatically turned off after
3 minutes.

1)

The State Command Data Field can have the following values:
0

Degas Off

1

Degas On

3.3.3.5 Emission Status

BPG400-SP

HPG400-SP

3.3.3.6 Emission Current

(BPG400-SP Only)

3.3.3.7 Degas Status

(BPG400-SP Only)

3.3.3.8 Hot Cathode Block State

Command
(BPG400-SP Only)