tira36e1-a (0310) BPG/HPG400 v1.cp

31

Overrange is the highest valid measured value at which Reading Valid is still on
"1".

Pressure unit

Overrange

COUNTS

21 602.06

Torr

1.5×10

-2

mTorr (Micron)

1.5

mbar

2×10

-2

Pascal

2.0

Pressure unit

Overrange

COUNTS

24 333

Torr

7.5006168×10

-1

mTorr (Micron)

7.5006168×10

2

mbar

1

Pascal

100

Underrange is the lowest valid measured value at which Reading Valid is still on
"1".

Pressure unit

Underrange

COUNTS

6397.95

Torr

3.7503×10

-10

mTorr (Micron)

3.7503×10

-7

mbar

5.0×10

-10

Pascal

5.0×10

-8

Pressure unit

Underrange

COUNTS

8333

Torr

7.5006168×10

-7

mTorr (Micron)

7.5006168×10

-4

mbar

1×10

-6

Pascal

1×10

-4

3.2.3.9 Overrange

BGP400-SP

HGP400-SP

3.2.3.10 Underrange

BGP400-SP

HGP400-SP