Appendix c: electrical connections – INFICON HPG400-SP ATM to High-Vacuum Gauge (Profibus) User Manual

Page 41

Advertising
background image

tira36e1-a (0310) BPG/HPG400 v1.cp

41

Appendix C: Electrical Connections

Technical data of gauges

→ [1],[2] [3] for BPG400-SP and [4], [5], [6] for

HPG400-SP.

Electrical connection
Pin 1 Relay Switching function A, common
Pin 2 Signal output (measuring signal)

0 … +10 V

Pin 3 Threshold value (Setpoint) A

0 … +10 V

Pin 4 Relay Switching function A, normally open contact
Pin 5 Supply common, GND
Pin 6 Threshold value (Setpoint) B

0 … +10 V

Pin 7 Degas on, active high

1)

+24

V

Pin 8 Supply of electronics unit

+24 V

Pin 9 Relay Switching function B, Common
Pin 10 Gauge identification:

BPG400-SP:

R

Id

= 42 kOhm

HPG400-SP:

R

Id

= 56 kOhm

Pin 11 Relay Switching function B, normally open contact
Pin 12 Signal common GND
Pin 13 RS232, TxD
Pin 14 RS232, RxD
Pin 15 Shielding, housing GND

BPG400-SP, HPG400-SP

TxD

RxD

Degas

1)

+U

b

R

Id

SP A

SP B

Measuring

signal

Threshold value, SP A
Threshold value, SP B

3
6
1
4

9

11

13
14

7
8
2

12

5

15

1.25 AT

24V

Degas

Identification

RS232

10

8

9

1

15

D-Sub, 15 pins

female,

soldering side

1)

Degas function only for BPG400-SP; pin 7 is not connected in the HPG400-SP.

Sensor cable connection
BPG400-SP, HPG400-SP

Advertising
This manual is related to the following products: