JULABO MA-26 Heating Circulators User Manual

Page 24

Advertising
background image

Setting of temperatures

24

7.

Setting of temperatures

The function of the

key is configurable.

1. If the key is pressed, normally only one adjustable working

temperature is displayed (factory setting).

2. Using the Menu Configuration which is started by pressing the

key a menu with three pre-set setpoints can be

assigned to the

key.

Press

key if a value is to be retained.

7.1.

1-setpoint mode / Direct setting of temperatures

The circulator uses the setpoint of t1 or t2 or t3 for temperature
control.

The indicated setpoint temperature can be changed directly any
time.

Example: change 25.00 °C to 50.00 °C

1. By pressing the key

the circulator switches to the

active >Setpoint< in the example on the left > t1 25.00°C<.
The integer digits flash

(example: <25>).

2. Change the value by pressing the keys

and

to

50.00 °C and confirm by pressing the key

.

The decimal digits flash and can be adjusted if desired.
Confirm once more by pressing the

key.

The end of the adjustment is signalled by the flashing message
>t1<

If the

„Up / Down“ keys

are pressed immediately

instead of pressing the key

this is called d

irect

temperature setting.

The circulator uses the new working temperature value for
temperature control.

The temperatures can be set in start or stop mode.

7.2.

Using the pre-setting in the

menu

Factory setting:

25 °C

37 °C

70 °C

Press the

key to call up the menu for temperature setting.

3 different working temperatures can be adjusted. Their values are
freely adjustable with the working temperature range.

Important:

Prior to the adjustment switch-over to the 3-temperature mode has
to be effected in the menu configuration.