9.1.

MENU PUMP – Setting of pump pressure

Factory setting:
stage

1

The capacity of the circulating pump is set by adjusting the motor
speed

Settings: stage 1 ... 4

Flow rate:

11 ... 16 l/m

Pump pressure:

0,22 ... 0,45 bar

1. Press the

key. Menu >Pu< is indicated.

2. Press the

key. The set parameter flashes (example: > 2<)

3. Change the parameter by pressing

or

and confirm by

pressing the

key.

or
Press the

key if the parameter is to be retained.

The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.

 Using a screwdriver turn the screw (1) anti-clockwise by

360 °.

 Using flat pliers turn the marking of the slide (2) to the

desired position.

 Tighten the screw.

Examples:

Internal applications in the bath

A 100 % internal bath circulation

(for large bath tanks)

B Reduced internal bath circulation

(for smooth surface of bath fluid)

External/internal applications

C 40 % external discharge,

60 % internal circulation
(for large bath tanks)

D 80 % external discharge,

20 % internal circulation
(for small bath tanks)