Safety installations, warning functions

28

8.2.

Switch-over from warning to shutdown function

If a shutdown of functional elements (e.g. heater, circulating pump) is
required when the limit values are exceeded or undercut the
circulator can be changed over from warning function >WARNING<
to shutdown function >ALARM<.

Factory setting:
>0 = WARNING<

1. Press the key

.

2. Select the menu >Aty (ALARM-TYPE)< by pressing the

key.

3. Press the

key and the set parameter will flash

.

(Example: 0)

4. Change the parameter by pressing the

key and confirm by

pressing the

key.

or
press the

key if the parameter is to retained.



Setting >0 = WARNING<
A mere warning function with optical and audible warning signal
(interval tone) A message appears on the MULTI-DISPLAY
(LED):

or

OVERTMP

SUBTEMP

 Setting >1 = ALARM<

Temperature limit with shutdown of heater and circulating pump.
An audible alarm sounds (continuous tone) and a message
appears on the MULTI-DISPLAY (LED):

or

OVERTMP

SUBTEMP