9.5.

MENU ATC - Absolute Temperature Calibration

ATC serves to compensate a temperature difference that might occur
between circulator and a defined measuring point in the bath tank
because of physical properties.


Example:
1-point calibration

°C

°C

T 1

T


3-point calibration

°C

°C

T 1

T

T

T

1 = Original curve

Principle:
For ATC calibration, in steady state the bath temperature at the
location of the temperature sensor (CT) is determined at the
respective adjusted working temperature. This value is then set on
the circulator in the menu >ATCalibration< under menu item >Ctx<.
This can be a 1-point, 2-point or 3-point calibration.

°C

TT

M

B (INT)

T

M = Temperature measuring instrument with temperature sensor
B = Bath tank (INTernal or EXTernal)
T = circulator

CT = Temperature on measuring point
TT = Temperature on circulator

Menu level 1

Level 2

Parameter level

Press the

key if parameter is to be

retained. Correction function for parameters or
values (prior to OK).

or

>StA< - ATC Status

The parameter flashes, switch by pressing

and

>NO< Carry out an ATC calibration

>YES< return to standard operation after
calibration.