JULABO MA-26 Heating Circulators User Manual

Page 29

Advertising
background image

MA

29

8.3.

Over and Sub temperature warning function

Factory settings:

Over temperature

t High

205 °C

Sub temperature

t Low

-55 °C

If the observance of a working temperature value >t 3< has to be
supervised for a sensitive temperature application, then set over
and sub temperature warning values. In the example below the
setpoint >t 3< 85 °C is surrounded by the values >t High< 87 °C
and >t Low< 83 °C. The electronics immediately register if the
actual temperature breaches one of the set limit values. The
resulting reaction is defined in the menu item >Aty (ALARM-
TYPE)< refer to (page 28).

1. Press the key

.

2. By pressing the

or

key select the menu > tHigh <

or > tLow <.

3. Press the

key. The integer digits flash.

4. Change the values to 87. °C and/or 83. °C by pressing the

and

key and confirm with the

key.

The decimal digits flash and can be adjusted if desired.
Confirm once more by pressing the

key.

See above examples.

The warning functions are only activated if the actual bath
temperature remains within the set limit values for 3 seconds
after switch-on.

Recommendation:

Set the over temperature warning value > t High < 5 °C to 10 °C
above the working temperature setpoint.

Set the sub temperature warning value > t Low < 5 °C to 10 °C
below the working temperature setpoint.