JULABO MA-26 Heating Circulators User Manual

Page 45

Advertising
background image

MA

45

11. Electrical

connections

Notice:

Use shielded cables only. The shield of the connecting cable is electrically
connected to the plug housing.
The unit ensures safe operation if connecting cables with a maximum length of
3 m are used. The use of longer cables does not affect proper performance of
the unit, however external interferences may have a negative impact on safe
operation (e.g. cellular phones).

/ Control output

The

connector may be used for control of JULABO refrigerated

circulators or as output for alarm messages.

1

3

2

4

5

Circuit:

Operation

= relay powered

Alarm

= relay not powered

Pin assignment:
Pin

Signal

1

+24 V (I max. current 25 mA)

2

0 V

3

Alarm relay

4

Reserved - do not use!

5 Cooling

pulse

RS232 serial interface

This port can be used to connect a computer with an RS232 cable
for remote control of the circulator.

9

6

5

1

RS232C

Pin assignments RS232:

Pin

2

RxD

Receive

Data

Pin

3

TxD

Transmit

Data

Pin 5

0 V

Signal GND

Pin 7

RTS

Request to send

Pin 8

CTS

Clear to send

Pin 1; 4; 6, 9 Reserved - do not use!

RS232 interface cable
Circulator

(9-pol)

PC

(9-pol)

Pin 2 RxD

Pin 3 TxD

Pin 3 TxD

Pin 2 RxD

Pin 5 GND

Pin 5 GND

Pin 7 RTS

Pin 8 CTS

Pin 8 CTS

Pin 7 RTS

Accessories: Order

No.

Description

8 980 073

RS232 interface cable 9-pol./9-pol. , 2,5 m

8 900 110

USB interface adapter cable