JULABO MA-26 Heating Circulators User Manual

Page 34

Advertising
background image

Menu functions

34

Warning

:

For supervised or unsupervised operation with the “AUTOSTART“ function
avoid any hazardous situation to persons or property

Take care to fully observe the safety and warning functions of the circulator.


9.2.4. Reset - Factory setting

>YES< resets all values to factory setting.

A >rSt< RESET can be effected in the >

< mode only.

Switch off the circulator by pressing the key

and call up the

menu CONFIGURATION.

During the message – init - all parameters are reset to factory
settings







9.3.

MENU Control parameters – Xp, Tv, Tn

Menu level 1

In most cases the control parameters preset in the factory are adequate
for achieving an optimum temperature sequence.
The control parameters allow adjustment to special control processes..

Level 2

Parameter level

Press the

key if a parameter is to be

retained. Correction function for parameters or
values (prior to OK)

0.1 … 99.9

Proportional range >Xp<

The parameter flashes, switch by pressing

and

3 … 9999

Reset time >Tn< (Integral component)

The parameter flashes, switch by pressing

and

0 … 999

Lead time >Tv< (Differential component)

The parameter flashes, switch by pressing

and