Troubleshooting guide / error messages

44

Error in A/D converter

Excess temperature protector defect.

The protection temperature is below the set working temperature
setpoint.
Set the protection temperature to a higher value.

The cable of the excess temperature protector has been disconnected or
short-circuited

The early warning system for low level reports a critical fluid level. Refill
bath fluid.

60

30

15

45

4 s

By quickly switching off and restarting the unit the alarm is

cancelled.
If the error occurs once more after the restart, a remote diagnosis is
required.


Configuration Error“
The configuration of the circulator does not correspond with its current
application.
Press the

key for a non-recurring, automatic change of the

configuration.
In this case please call the JULABO Technical Service or an authorized
dealer.


Disturbances that are not indicated.

The electronic pump motor is overload-protected by an electronic current
limiter. If viscosity of the bath fluid is or becomes too high, the motor
stops running.

15

A
M
P

Mains circuit breakers (resettable) 15 A