JULABO MA-26 Heating Circulators User Manual

Page 38

Advertising
background image

Menu functions

38

>tYP< - ATC TYPE

The parameter flashes, switch by pressing

and

A >1-point<, >2-point< or >3-point<
calibration
can be carried out.

The value > tt1< is only indicated

In addition the measured temperature value
>Ct 1< is saved during the next step.

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

If only a 1-point calibration is carried out, the following menu items
are not indicated anymore

The value > tt2< is only indicated

In addition the measured temperature value
>Ct 2< is saved during the next step.

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

If only a 2-point calibration is carried out, the following menu items
are not indicated anymore

The value > tt3< is only indicated

In addition the measured temperature value
>Ct 3< is saved during the next step.

Integer digits flash, set by pressing

+

Decimal digits flash, set by pressing

+

Advertising