9.6.

MENU LIMITS

Menu level 1

>Li< = LIMITS – limitation of operating temperature range

Maximum and minimum setpoint
Restriction of the adjustable temperature range.

Level 2

Parameter level

Press the

key if parameter is to be retained.

Correction function for parameters or values (prior
to OK).

 Integer digits flash, set by pressing

+

 Decimal digits flash, set by pressing

+

 Integer digits flash, set by pressing

+

 Decimal digits flash, set by pressing

+

Factory settings::

200

°C

(Setpiont High)

-50

°C

(Setpoint Low)

The limitation of the operating temperature range effects the

temperature setting in the menu with the key

.

Only setting of working temperatures which lie within the
determined limits is possible

This applies to settings in the MENU

,

(refer to page 24)

and for settings in the MENU

high temperature

low temperature

(refer to page 29)

The temperature values are automatically deferred into the limit
range.

Setting range: -50,0 °C ... +200,0 °C