KD Scientific Legato 200 Series User Manual

Page 11

Advertising
background image

11

Product Over view

K

D

S

L

e

g

a

to

2

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Easy external connections to a computer or other control

devices are made through USB interface or RS232 (9 pin

Dsub). Simple ASCII commands make communication

with the pump easy. For direct control of the pump, the

user can use the I/O interface (15pin Dsub). Two trigger

outputs and a trigger input in for external events such as

a process parameter is met. The footswitch input will

allow the control of the pump through an external device.

The unit also has an output for run indication allowing

connection to a remote light.

All the models accommodate dual syringes except for the

push/pull pumps which have two syringes on each side

for simultaneous infusion and withdrawal. Complete

flexibility is offered with the user able to use most

manufacturer’s syringes with sizes ranging from 0.5 ul to

140 ml. Modular configuration is now done through the

field installable syringe mechanisms. These optional

accessories provide the necessary brackets to hold up to

four 140 ml syringes or six 10 ul syringes. Also, the user

can use the microliter rack which has been specifically

designed for smaller syringes.

The programmable models offer maximum flexibility and

capability for configuring and running different programs.

Up to 40 programs of 20 steps each can be configured

and stored in the unit, and recalled quickly with the touch

of a button. The programs are identified with a 15

character alphanumeric name for easy identification and

can be stored on the computer and downloaded at future

dates. A real time clock is available to start and stop

programs based on time. Extensive programming options

allow the user to duplicate configurations, append one

configuration to another, rename programs, delete

programs, export programs to a PC or export a program

to another pump. This feature can ensure consistency and

accuracy of multiple pumps.

Preprogrammed profiles are selectable from Constant

Rate, Ramp, Stepped and Pulse. These standard profiles

make configuration easy to set up. If more complexity is

needed the user can select from advanced programming

options including:

Go To statements

Time delays

Repeating steps

Linking different programs

Stopping the pump

Triggering the pumps TTL output

Accepting an event input in such as a user

touch, motor stall

Events can trigger the pump to withdraw or infuse

The pump Information/setup screen shows the pump’s

software version, calibration and lubrication intervals. The

pump’s alarm configurability includes alarms for near

end of run (user selectable), completion of run, power up,

keypad clicks, stall detection and calibration reminder.

Two options are available for the Legato 200 Series. The

analog input option which allows the analog control of the

motor speed. By applying a 10 VDC max to the circuit the

motor speed can be varied. The second option is for an

internal fan. These will be factory installed.

Pump Models

The Legato 200 Series is available in 5 different

configurations.

Legato 200

Infuse only syringe pump. Accommodates

2 syringes. User definable flow rates with selectable target

volumes or time values to control the total infusion volume.

Legato 210

Infuse/withdraw syringe pump.

Accommodates 2 syringes. This unit supports infuse only,

withdraw only, infuse/withdraw, withdraw/infuse and

continuous mode. User defined flow rates with selectable

target volumes. One Program can be stored.

Legato 210P

This is a Programmable Legato 210 with

user defined configurations/programs of up to 800 steps.

Up to 40 programs of 20 steps each can be stored in

memory.

Legato 270

Push/Pull syringe pump. Accommodates

2 syringes for infusion and 2 syringes for withdrawal.

This model supports infusion and withdrawal simultaneously

at user defined flow rates and with selectable target

volumes to control the total volume pumped, It also

supports infuse only, withdraw only, infuse/withdraw,

withdraw/infuse and continuous mode. User defined flow

rates with selectable target volumes.

Legato 270P

This is a Programmable Legato 270,

Push Pull pump with user defined program configurations of

up to 800 steps. Multiple programs can be stored in memory.

Modular syringe racks can be purchased to create a

multichannel syringe pump .

10 channel syringe rack

4 channel syringe rack

µl syringe rack

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

Advertising