KD Scientific Legato 200 Series User Manual

Page 27

Advertising
background image

27

Configuring the KDS Legato 200 Series

K

D

S

L

e

g

a

to

2

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Setting the Pump Address

Up to 100 pumps can be linked together. Each pump

can be given a unique address.

In the Set-up Screen press

Pump Address

.

Valid entries are 00 to 99.

Enter the pump address.

Accept

the changes made by pressing

and go back to the Set-up screen.

Cancel

the changes made by pressing

and go back to the Set-up screen.

Press Here

Press Here

Press here to

select Baud Rate

Setting the RS 232 Baud Rate

The Legato 200 Series has an RS 232 port for serial

communications.

To adjust the baud rate from

9600 to115,200: in the Set-up screen

press

Baud Rate

Select the Baud Rate.

Accept

the changes made by pressing

and go back to the Set-up screen.

Cancel

the changes made by pressing

and go back to the Set-up screen.