KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

5. Screw down the syringe clamp knob (6) until the

syringe barrel clamp (10) is tight against the

syringe barrel. Do not overtighten, especially

when using glass syringes.

6. Secure the syringe flanges by pushing the retaining

brackets (4,5) tightly against the syringe flanges

using the bracket clamping knobs (7,8). The

syringe plunger should be secured by the pusher

block retaining bracket and the flange of the

syringe should be secured by the syringe holder

block retaining bracket.

7.

Tighten the mechanical stop collars (9) on the

pump to prevent excess pusher block travel.

NOTE: The syringe barrel clamp (10)

may need to be flipped depending on your

syringe size. To do so, completely unscrew

and remove the syringe barrel clamp knob

from the syringe block. Then remove lock

nut (11). Flip the bracket and reassemble.

SYRINGE LOADING

2

5

4

1

3

9

7

8

10

6

> 30 cc syringes

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: For infuse/withdraw models make sure

the plunger is secured in the retaining bracket

and the barrel of the syringe is secured in the

retaining bracket.

The Standard Legato 200 Series can accommodate up to two

syringes. Any type of syringe, glass, plastic or stainless steel

can be used. The pump can accommodate from 0.5 ul to 140

ml syringes. The syringe barrel clamp should be flipped for

syringes larger than 30 ml. For syringes smaller than 30 ml,

the syringe clamp should be used as shown.

1.

Squeeze the release tabs (1) and move the pusher

block (2) to the approximate length of the syringe.

2. Unscrew the bracket clamping knobs (7,8) to

loosen the retaining brackets (4,5) on the pusher

block (2) and the Syringe holder block (3).

3. Unscrew the syringe barrel clamp by unscrewing

the syringe barrel clamp knob (6).

4. Place the syringe on the syringe holder block (3).

Ensure the barrel flange is with in the syringe holder’s

block’s retaining bracket (5) and the plunger flange is

in the pusher’ blocks retaining bracket.

< 30 cc syringes

11

!

CAUTION: Be sure mechanical stop

collars are positioned properly to

prevent the plunger from bottoming

out on glass syringes.