PUMP CHAIN COMMANDS

The Pump Chain commands allow all pump control

information to be managed from an external computer

source. These commands can also be used to control a

series of pumps (up to 100) from a single computer

interface.

KDS Legato 200 Series commands are communicated to

the pump via the RS-232 or USB port interfaces through

a terminal program such as HyperTerminal. In using the

Pump Chain commands, you will need to assign each

pump in the pump chain a unique address, using the Set

Pump Address command in the Settings menu. The address

range is from 00 to 99. This address value is used to

identify which pump is to receive a command and which

pump is responding. The first pump in the chain, the one

connected to the computer, must use address 0 (zero).

Configure each pump with its assigned address and baud

rate as described in the Pump Settings section of this guide.

Using the KDS Legato 200 Series

Pump Chain Commands

Commands may be abbreviated to the first four letters,

i.e. address would be abbreviated addr. A space must

follow the command if arguments are included.

If the pump address is nonzero, the one or two-digit pump

address precedes the command without any separator.

For example, to set the infuse rate for pump 12, the

command would look like “12irat 3.2 u/m”. If the pump

address is one digit, it doesn’t have to be preceded by a 0.

NOTE: If the USB port is used on a pump

with a nonzero address, the pump address

does not need to be used.

NOTE: When daisy-chaining using the RJ-11

connectors, all pumps must use the same

baud rate. Values above 38400 are

recommended.

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

NOTE: Once communication is established,

if you manipulate the touchscreen, you must

reestablish communications to the PC

(<CR> will do this)

NOTE: System commands and start/stop

commands can be executed from program

or mode configuration screens. For

parameter commands (rate, dia, vol, etc) it

is necessary to execute from the

Configuration Run screen.

Every line of the response from a pump with a nonzero

address is prefixed by the two digit pump address (prefixed

with a 0 if necessary) followed by a colon. No colon is

used to separate the pump address from the prompt however.

In the command list below, the following convention is used:

{ }

Required parameter

[ ]

Optional parameter

|

Separator between parameter choices

#

Numeric value without preceding zeros

###

Numeric value with preceding zeros

#-#

A range of values

<cr>

Carriage return

<lf>

Line feed

<sp>

Space

[prefix]

Pump address prefix in the format #: if the

pump address is not zero

<prompt>

Prompt (see below)

The following prompts are returned after a command is

executed:

:

The pump is idle

>

The pump is infusing

<

The pump is withdrawing

*

The pump stalled

T*

The target was reached

If the pump is in poll mode, an XON character is added

after the prompt.

If the pump address is nonzero, the pump address is

prefixed to the prompt without a colon.

In the following list of commands:

The pump address is prefixed to every response line

followed by a colon.

<lf>[##]<prompt> is the response unless otherwise noted.

Error messages

Error messages are displayed if the entered command

cannot be executed for some reason. The error message

will take up two lines with the first line being the message

type and the second line describing the error itself. The

second line may be up to 80 characters long.

NOTE: To maximize communication speed

[as fast as 50 ms rate changes]: (a) prefix

commands with the @ symbol to turn off GUI

updates (ex: @irate 100 u/m) (b) use the

‘NVRAM off’ command to turn off writes of

rate to memory.