Upgrading Legato 200 Series Software

1. Upload the latest software version to your desktop (format is filename.srec).

2. Disconnect all I/O devices and then connect the pump to PC using a USB cable.

3. On the Set up screen of the pump, press the Upgrade Software button twice to enter “Boot Loader

Mode”.

4. From the CD provided with the pump, open the Legato 200 Series Pump Updater Application.

5. Click Open Update File and browse to the file previously saved on your desktop and click Open.

6. Click Start Update.

7. After the update is complete, press Reboot to power cycle the pump.

NOTE: Prior to upgrading, users will need

to install the KDS Legato 200 Series USB

driver as well as the bootloader driver file

(KDS Legato 200 Series-Bootloader.inf).

Reference “USB Virtual Comport Driver

Installation”.

NOTE: To exit boot loader mode without

upgrading the software, power cycle the

pump.