KD Scientific Legato 200 Series User Manual

Page 78

Advertising
background image

78

Appendices

K

D

S

L

e

g

a

to

2

0

0

S

e

ri

e

s

U

s

e

r’s

M
a

n

u

a

l

KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Syringe Loading – Push-Pull Mechanism

phone

508.429.6809

fax

508.893.0160

e-mail

[email protected]

web

www.kdscientific.com

Installing a Syringe

1.

Loosen Thumbscrews (1) to release the Retaining

Brackets (2 & 3) located on the Floating Syringe

Holder (4) and Pusher Block (5).

2. Loosen Syringe Clamp (6) by unscrewing the

Syringe Clamp Knob (7).

3. Place the syringe(s) under the Syringe Clamp on

the Floating Syringe Holder. Ensure that the

syringe’s flanges are within the Retaining Brackets.

4. Lock the flanges into place by tightening the

Retaining Brackets using the Thumbscrews.

5. Tighten the Syringe Clamp using the Syringe

Clamp Knob.

Repositioning the bracket location

1.

Unscrew the syringe clamp completely

2. Push the knob out of the clamp

3. Reposition and/or flip over and then push the knob

back into the clamp

1.

Thumbscrews

2.

Retaining Bracket,

barrel flange

3.

Retaining Bracket,

plunger flange

4.

Floating Syringe holder

5.

Pusher Block

6.

Syringe Clamp

7.

Syringe Clamp Knob

8.

Locking pins

9.

Stationary Syringe holder

1

4

8

9

6

7

2

3

5

Repositioning the Adjustable Syringe Holder

[For syringes > 60 ml]

For larger syringes you will need to detach the Floating

Syringe Holder and then mount it on the outside of the

Stationary Syringe Holder.

1.

Loosen the two Thumbscrews (1) on the Stationary

Syringe Holder (9) and slide the Retaining Bracket

(2) forward.

2. Remove the two Locking Pins (8)

3. Remove the Floating Syringe Holder (4).

4. Place the Floating Syringe Holder on the outside

of the Stationary Syringe Holder and lock into

place using the Locking Pins.

5. Slide the Retaining Bracket back and tighten using

the Thumbscrews.

Locking Pins