MTS SWIFT 30 Sensor User Manual

Page 129

Advertising
background image

SWIFT 30 Sensors

Troubleshooting

129

Spinning Zero Procedure:
The Bridge Zero and Angle
Zero indicators stay on too
long, or they continue to
blink slowly, even after the
wheel has rotated twice.

The angular output signal is

not reaching the TI. The TI

will wait for the two
revolutions as noted by the

encoder. If the encoder

signal is not reaching the TI,
it may eventually time out

and blink slowly, indicating

a zero failure.

Check that all cables are attached and

undamaged. In particular, check the main signal

cable from the transducer to the TI.

Check that the encoder output is present. The

angle output signal from the TI should be a 0–5
V sawtooth output while the transducer is

spinning at a constant velocity.

The EncoderSize parameter

in the calibration file is set

incorrectly.

Check that the EncoderSize parameter is set

properly for the slip ring (EncoderSize=1).

The indicators will stay lit if

there is no encoder signal. If

the wheel stops rotating or
the encoder is disconnected

after data collection has

started, the data collection
will time out and blink

slowly.

If this signal is not present, check that the slip

ring/encoder assembly has not been damaged. If

you suspect encoder damage, swap the slip ring
assembly with a known functioning unit to

verify it. If the slip ring/encoder assembly is

damaged, call MTS or replace it with a spare
slip ring assembly.

Some or all transducer
output signals read 0 volts
at the TI output even when
a load is present.

The TI is not turned on.

Check that TI is turned on and the green power
indicator is lit.

The transducer signal cable

is not connected.

Check that the cable is connected between the

back of the TI box and the slip ring (or
connector housing for a road simulator).

Output cables are not

connected.

Check that the output connectors are securely

fastened to the data acquisition device.

Cables or connectors are

damaged.

Check all cables and connectors, particularly the

power and signal cable. Cables must be free of
nicks or cuts, and all connector pins shown on

the cable drawings must be present and not bent.

The shunt calibration connectors intentionally

have some pins missing.

Data acquisition
configuration, cabling, or

input ports are incorrect or

damaged.

Verify that the data acquisition is configured
correctly. While the TI is turned on and all

cables from the SWIFT transducer to TI are

connected, use a handheld DVM to verify
output on individual channels while they are

being loaded. (Use the BNC connectors located

on the back panel of the TI.)

Grounding problem.

Check that there is a good signal-to-ground

connection between the TI black lug, the signal

ground, and the ground of the external data
acquisition equipment.

Troubleshooting Guide (part 2 of 12)

S

YMPTOM

P

OSSIBLE

C

AUSES

S

OLUTION

Advertising