MTS SWIFT 30 Sensor User Manual

Page 133

Advertising
background image

SWIFT 30 Sensors

Troubleshooting

133

The data acquisition scales

are set incorrectly.

Check that the data acquisition scales are set

correctly. The full scale calibration range is

shown on the Calibration Report, with 10 V =
Full Scale (unless you requested custom

calibration output before calibration at MTS).

With the TI turned on and all cables from the

SWIFT transducer to TI connected, use a

handheld DVM to check the output on
individual channels while they are being loaded

to make sure that the output at the TI box and

data acquisition correlate. (Use the BNC
connectors located at the back panel of the TI.)

Output levels are much
higher or lower than
expected. (Continued)

The transducer was zeroed

with load applied (or a
different load than the

intended tare weight for

non-spinning applications
only).

Rezero the TI, being careful not to touch or load

the transducer during the zeroing procedure.

A zero offset is present in

the data.

See Zero Offset on the previous pages of this

Troubleshooting Guide.

Shunt Calibration indicator
light blinks after shunt
calibration (Shunt
Calibration Failed)

The shunt cables are not

connected, or are damaged,

or their connectors are not
secured.

Check that the shunt connectors were securely

installed during the shunt calibration procedure.

Verify that the cables are not nicked, cut, or
damaged, and that no connector pins are

damaged. Make sure that pins shown in the

cable drawings are securely positioned in the
connector.

The signal cable is not

connected to the SWIFT
transducer.

During a shunt calibration, the shunt cables

provide the shunt across the individual bridges,
but the outputs that are verified in the TI come

from the output signal cable. Check that the
output signal cable is not damaged and is

securely fastened to both the TI and the

transducer.

Troubleshooting Guide (part 6 of 12)

S

YMPTOM

P

OSSIBLE

C

AUSES

S

OLUTION

Advertising