Tishunt - transducer interface shunt, Tishunt - transducer interface shunt 50 – MTS SWIFT 30 Sensor User Manual

Page 50

Advertising
background image

SWIFT 30 Sensors

50

TISHUNT - Transducer Interface Shunt

Software Utilities

TISHUNT - Transducer Interface Shunt

This program is a utility with various functions related to shunts. The SWIFT
system includes the ability to connect a shunt resistor across each of the resistive
bridges in the transducer. This shunt function can be used as a simple verification
that the SWIFT system is working normally. Shunt verification activates the
shunts and compares the results to those recorded during calibration. While this
does not guarantee the transducer is still in calibration, it provides some level of
confidence it is working normally. If the shunt results differ significantly from
those recorded during calibration the SWIFT system should be evaluated for
possible problems.

Syntax

The tishunt command requires one argument, as follows:

tishunt port box

port is the communications

1 = COM1:

2 = COM2:

box is the SWIFT-TI communications address

1 = lowest possible address

9 = highest possible address

Example

The Shunt Main Menu options are described in the following paragraphs.

Initializing box...SWIFT TI Shunt Main Menu

0...exit
1...Read current Shunt values
2...Set the TI shunt tolerance
3...Restore Shunt parameters from a file
4...Scan inputs & outputs with shunts
5...Command a shunt cal
6...Write last shunt measurements to a parameter file

Enter choice:

Advertising