MTS SWIFT 30 Sensor User Manual

Page 134

Advertising
background image

SWIFT 30 Sensors

134

Troubleshooting

Excessive noise on the

signal caused the shunt

tolerance to be exceeded.

Electrical Noise: The SWIFT sensor has power

conditioning, shielded cables, and on-board

amplification to reduce electrical noise.
However, if the TI or transducer are very near a

powerful noise source, some noise can be

picked up with the signal.

Check that the TI electronics are grounded

properly at the power cabling or grounding bar
located on the back panel of the TI box. If

possible, perform the zeroing and shunt

calibration procedure away from the powerful
noise source.

Physical Noise: If the transducer is part of an
assembly that is connected to active hydraulics

or other assemblies that may experience

physical vibrations, these vibrations may be
picked up as inertial forces read by the

transducer. Check that the transducer is not

experiencing dynamic loading or vibrations
during the zero process or shunt calibration.

A temporary fix is to increase the shunt
verification tolerance by changing the setting in

the TI. Note that any temporary fix by

increasing the tolerance may allow a shunt
calibration to pass, when it may be indicative of

a problem that should not be ignored.

Shunt Calibration indicator
light blinks after shunt
calibration (Shunt
Calibration Failed).
(Continued)

The slip ring or road
simulator connector housing

assembly is not transmitting

a signal.

The shunt procedure requires the output through
the slip ring or connector housing. Make sure

that the slip ring/connector housing connection

is secure and that no connectors appear to be
damaged. If possible, swap it out with another

slip ring or connector housing to verify that the
slip ring or connector housing component has

not failed or been damaged.

The shunt reference values
were changed.

Use the TIXFER utility to upload the current
calibration file. Check the file to verify that the

variables ShuntDeltaRef are the same as shown

on the original calibration file provided by
MTS. The eight ShuntDeltaMeas values

should read approximately -3.9 V or -2.8 V for

both the aluminum and titanium standard
SWIFT transducers.

The shunt tolerance is not

set correctly.

Use the TIXFER utility to upload the current

calibration file. The ShuntTolerance value
should be set at 2, indicating an allowable shunt

variation of 2%. If this value is changed to zero

or a very small number, the difference between
the reference and the current shunt

measurement may fall outside of this tolerance

and be read as a shunt failure.

Troubleshooting Guide (part 7 of 12)

S

YMPTOM

P

OSSIBLE

C

AUSES

S

OLUTION

Advertising