Edit the Calibration File

SWIFT 30 Sensors

Setting up the Transducer Interface

67

2. If necessary, edit the value for OutputPolarities.

This value sets the polarity states of all six outputs (Mz, My, Mx, Fz, Fy, and
Fx) using the following binary scheme (the default configuration is
represented):

For example, if you want to invert Fx and Fy, and Mx and My, (making their
polarities opposite those shown on the labeling) you would make the
following changes:

O

UTPUTS

M

Z

M

Y

M

X

F

Z

F

Y

F

X

B

IT

2

5

2

4

2

3

2

2

2

1

2

0

B

IT

V

ALUE

32

16

8

4

2

1

E

XAMPLE

1

0

1

0

0

0

= 40

10

(default)

M

Z

M

Y

M

X

F

Z

F

Y

F

X

B

IT

2

5

2

4

2

3

2

2

2

1

2

0

B

IT

V

ALUE

32

16

8

4

2

1

1

0

1

0

0

0

= 40

10

(old value)

1

1

0

0

1

1

= 51

10

(new value)

*

* See the examples in the next table.