Routine maintenance overview checklist – MTS Landmark System User Manual

Page 110

Advertising
background image

Routine Maintenance Overview Checklist

Recommended service to be performed at each running time interval noted

Annually

Biweekly

Weekly

Daily

Calendar Time using 8 hour Running
Time rate per day

2,000

1,500

1000

500

80

40

8

Running Time-Hours

X

1

Check that Actuator Platen Area is Clean

X

Monitor Filter Indicators

X

Check Hoses/Cables/Connectors

X

Check Crosshead/Lifts/Supports

X

Check Actuator to be Dry

X

Check Hydraulic Service Manifold

X

Check Lift Seal Condition to be Dry

X

Check Lock Seal Condition to be Dry

X

Bionix Lubricate Axial/Torsional Spline
(75-100 hrs)

X

Check Actuator

X

Label Inspection

Crosshead/Load Frame

MTS

MTS

MTS

MTS

MTS

2

Check Crosshead/Lifts/Supports

MTS

MTS

MTS

MTS

MTS

Lift Seal Condition is Dry

1

Symbol denotes services performed by equipment operators. Most of these procedures involve visual
checks that should not interfere with test system operation. These checks are also completed by trained
field service engineers on each Routine Maintenance visit.

2

Symbol denotes service performed by trained field service engineers as part of an MTS Routine Maintenance
plan. Some of these procedures require special service tools and/or specific service training to complete.

110 | MTS Landmark Test System Operation

Maintenance

Advertising