Zero the force signal – MTS Landmark System User Manual

Page 68

Advertising
background image

For specimen installation, select a low compression limit to reduce the chance of specimen damage
or injury. While -0.05 kN is shown here as an example, you should use a limit that adheres to your lab's
specimen installation guidelines.

a) Enter -.050 kN in the Lower Limit text box (a negative number indicates compression). Be sure to

press Enter after typing in the limit.

b) Once again, select Interlock for the Lower Action.

Zero the Force Signal

1. Select Exclusive Control.

In the Station Manager application main window, select the Exclusive Control check box. This allows
control to come from either the computer software or the handset, but not both.

2. Zero the force signal.

68 | MTS Landmark Test System Operation

Running the Example HCF Test

Advertising