Adjust limits for the test – MTS Landmark System User Manual

Page 89

Advertising
background image

a) If you find the values in the Meters window changing at a rate less than the frequency of the Function

Generator, double the Sensitivity.

b) If you find the values changing at a rate faster than the frequency, halve the Sensitivity. Repeat this

process until you are satisfied with the results.

4. Switch Ch1 Force to a Peak/Valley Meter and adjust Sensitivity.

Use the same procedure to change the top axial force meter to Peak/Valley meter type.

Adjust Limits for the Test

Setting force limits just outside the expected values for the HCF test helps prevent specimen and equipment
damage if something unexpected happens during a test run.

1. You can retain the wide displacement limit range set earlier because this example tension test uses

no fixturing that requires protection.

a) From the Station Manager main window, select Display > Station Setup, and click the Channel

Input Signals button.

b) Select the Displacement channel.
c) Select the Limits tab.
d) Use the following parameters:

Ch1 Displacement (Upper Limit, Upper Action) — 100 mm, Interlock

Ch1 Displacement (Lower Limit, Lower Action) — (- 100) mm, Interlock

2. Adjust force limits so that they are just outside expected values for the HCF test.

MTS Landmark Test System Operation | 89

Running the Example HCF Test

Advertising