MTS Landmark System User Manual

Page 81

Advertising
background image

This introduces a force waveform to the specimen.

a) In the Station Manager main window, click the Function Generator button.
b) Set up a force wave form using the following parameters:

•

Channel — Ch1 (or other depending on your configuration)

•

Control Mode — Force

•

Command Type — Cyclic

•

Target Setpoint — +2 kN

•

Amplitude — 1 kN

•

Frequency — 1 Hz

•

Wave Shape — Ramp

•

Compensator — None

5. Open a Station Manager Scope.

Click the Scope button in the Station Manager main window.

6. Set up the Scope.

MTS Landmark Test System Operation | 81

Running the Example HCF Test