a) From the Station Manager main window, select Display > Station Setup, and click the Channel

Input Signals button.

b) Select the Force channel.
c) Select the Limits tab.
d) Use the following parameters:

Ch1 Force (Upper Limit, Upper Action) — 15 kN, Interlock

Ch1 Force (Lower Limit, Lower Action — (-1.00) kN, Interlock

Show Station Manager Scope and Meters for the Test

Show Station Manager Scope and Meters.

Note: Your meter readings will reflect your station and will be different from those shown.

90 | MTS Landmark Test System Operation

Running the Example HCF Test