20 • Downloading Data from the SC-01

3. Press and hold the POWER button on the SC-01 until you

hear a beep, then release it. The SC-01 will begin it’s power
up sequence.

4. If the CAL. LIMIT menu item in the SC-01’s Setup Mode is

on, then one of the two following startup sequences takes
place. If CAL. LIMIT is turned off, the startup sequence in
Step 5 below takes place. See the SC-01 Operator’s Manual
for a description of the CAL. LIMIT Setup Mode menu item.

If calibration is past due, the following screen appears with
“CAL.” flashing and the LEDs and buzzer pulsing several
times alerting you that the SC-01 is due for calibration. If
this screen appears, wait for the buzzer and LED pulsing
to stop, then press and release the POWER button to
continue in the startup sequence.

The SC-01 will then show the current date for a few
seconds before attempting to connect to the computer.

If the calibration is up to date, the SC-01 then displays the
calibration due date, how many days until the next

To Computer
Serial Port

To Computer
USB Port

USB Connector

IrDA/USB or
IrDA/Serial Cable

Cable's Infrared Port

Serial Connector

SC-01

Infrared Port

Figure 10: Aligning the SC-01 With the Cable Infrared Receiver