26 • Viewing, Printing, and Deleting Data in the Last Calibration Window

Viewing, Printing, and Deleting Data in the
Last Calibration Window

You can access data on the most recent successful calibration
for each instrument that has been downloaded in the Last
Calibration window. You can view, print (calibration date only),
and delete this data in the Last Calibration window.

Viewing and Printing Last Calibration Data

Open the Last Calibration window by clicking the Last
Calibration
button along the right side of the program window.
When you open the Last Calibration window the first time after
launching the program, it will open with the Need Calibration
view option selected. There are three view options in the Last
Calibration window: Need Calibration, Calibration Date, and
Calibration Record.

Need Calibration Option

Selecting this option shows the last calibration date and last
download date for the SC-01s that are due for calibration.

NOTE: The calibration interval value, the number of days after a

calibration that a new calibration is due, is saved in the
SC-01. If the calibration interval is changed in the SC-
01, the new calibration interval value will not be known
by the program until the SC-01 is downloaded using the
Complete Download button or the Instrument
Information
button.

NOTE: If an instrument’s calibration interval is turned off in the

CAL. LIMIT menu item of the SC-01’s Setup Mode (see
the SC-01 Operator’s Manual), the SC-01 Data Logger
Management Program will always show that calibration
is due for that instrument.