Rockwell Automation T7480 ICS Regent+Plus Guarded Analog Output Modules User Manual
Page 19
Guarded Analog Output Modules (T7480)
 
P D - 7 0 2 6 M a r - 0 6  
19
4. Connect the six precision resistors to the I/O extender
jumper posts as shown in Figure 11. Use test clips as 
required to make the connections. 
5. Remove the four screws on one side of the module and
remove the printed circuit board from the module 
clamshell housing. 
6. Install the printed circuit board into the I/O extender
module. Allow the board to warm up for approximately one 
minute. 
7. Connect the voltmeter's negative probe to test point TP4.
Measure the voltage at TP1. The voltage should be 
10.0000 VDC (±0.0003 VDC). Use R22 to adjust the 
voltage. 
8. Connect the voltmeter’s negative probe to TP4. Measure
the voltage at TP2. The voltage should be 8.0000 VDC 
(±0.0002 VDC). Use R17 to adjust the voltage. 
9. Connect the voltmeter negative probe to TP4. Measure
the voltage at TP3. The voltage should be 2.0000 VDC 
(±0.0001 VDC). Use R19 to adjust the voltage. 
10. Using the forcing functions of W
INTERPRET
, force all of the
six analog outputs to zero. Refer to the Regent User’s 
Guide for more details about the forcing functions. Wait 
approximately one minute after forcing the outputs to 
allow them to stabilize. 
11. Adjust each of the six Offset potentiometers until the
voltage measured across the corresponding precision 
resistor is +1.0000 V (± 0.0002 V). Connect the voltmeter 
across the resistor leads to measure the voltage. 
12. Force all six of the outputs to 4095. Wait approximately
one minute after forcing the outputs to allow them to 
stabilize. 
13. Adjust each of the six Gain potentiometers until the
voltage measured across the corresponding precision 
resistor is +4.9990 V (± 0.0002 V). 
After completing one iteration of the calibration, repeat the 
offset and gain adjustments (steps 10 through 13) until the 
readings are within specifications. It is necessary to repeat