9 stat key – Bowers Group Sylvac D100S Readouts User Manual

Page 23

Advertising
background image

22

1.6.9 STAT KEY

Allows two types of statistical results to be displayed:

Choice of display:

1/ press

2/ then

appears:

1 = display of classical statistics and bar chart

2 = display of

_

X/R control charts

These are the measured values, stored in memory using the Store key (refer to
Store key in Section 1.6.15), which are used. Each channel (1 to 64) can
have an independent statistic. The statistical results can also be given on the
Centronics and RS232 links in order to be printed or dealt with by computer.
The bar chart may be printed in 80 or 40 columns, while the

_

X/R cards may

be printed in 80 columns (refer to Print key in Section 1.6.14).

First choice: bar chart

1/ Press key

. Following display will appear:

N = number of stored values = used for statistical analysis

Xmax = maximum stored value

Xmin = minimum stored value

R = Range = Xmax - Xmin

_

X = mean value =

σ (sigma) = standard deviation =

s = standard deviation = standard deviation of population based on a sample

The

key allows statistical results to be printed via the Centronics or RS232 according to the selected format. It

is also possible to press Print in the following displays.

Statistical results may also be requested in scanning mode. The unit will first request the channel on which statistics are
required, and at the end request the next channel (if 0 is chosen as channel number, the statistics of all channels will
be printed).

2/ if the unit is in tolerance display mode, the next display can be requested with

Exit statistics mode with

or

.

CHANNEL 1
N

= 112

Xmax

= 10.007

Xmin

= 9.990

R

= 0.017

x

= 10.000

σ

= 0.0020

s

= 0.0020

Enter the choice :

1

:

Histogram

2 : X/R - chart

X

i

i=1

N

N

Advertising