4.5 TECHNICAL SPECIFICATIONS

Accuracy using extension cables:

These measuring errors are applicable only when using D100S or D101 units without recalibration:

Normal cable

+ extension up to 5 m

: additional error

1.5 µm approx.

Normal cable

+ extension up to 10 m

: additional error

3 µm approx.

Normal cable

+ extension up to 15 m

: additional error

6 µm approx.

Normal cable

+ extension up to 25 m

: additional error

8 µm approx.

Direct cable

up to 5 m

: additional error 3 µm approx.

Direct cable

up to 10 m

: additional error 6 µm approx.

Direct cable

up to 15 m

: additional error 10 µm approx.

This is a progressive error margin and recalibration of D100S unit can considerably reduce the error.

SPECIFICATIONS

CODE NUMBERS

Construction

Type of bearing for measuring plunger

LINEAR MEASURING RANGE

Total range

Accuracy over the measuring range

Increase of measuring force

Permissible lateral force

Repetability

Zero drift

Temperature range

Cable length

Extension cables

Direct cable special length

Lifting lever

Interchangeable measuring contact

0 - 50 deg. C

1 m 50 / 59''

up to 20 m / 787''

up to 20 m / 787''

M 2.5

P10 / P10L

P25

P50

Friction bearing

3.2 g / 0.11 oz 8.7 g / 0.31 oz 14.6 g / 0.51 oz

10 mm / 0.44'' 25 mm / 1'' 50 mm / 2''

10.8 mm / 0.43'' 25.8 mm / 1.02'' 52..2 mm / 2.06''

10.4 mm / 0.41'' 25.8 mm / 1.02'' 51 mm / 2.01''
0.4 - 0.5 mm 0.8 - 0.9 mm 1.0 - 1.2 mm

1 um /0.00004'' 1.5 um / 0.00006'' 4 um / 0.00016

0.6 - 0.9 N 0.6 - 1.2 N 0.6 - 1.4 N

0.03 N/mm 0.024 N/mm 0.016 N/mm

0.6 N 0.3 N 0.25 N

0.2 um/0.000008'' 0.2 um/0.000008'' 0.4 um/0.000016''

0.02 um/deg.Cmm 0.01 um/deg.Cmm 0.01 um/deg.Cmm

Friction bearing

Friction bearing

Plunger gage

Plunger gage

Plunger gage

P5(L) / P5(L) V

Friction bearing

Plunger gage

3.2 g / 0.11 oz

5 mm / 0.2''

6.5 mm / 0.26''

5.7 mm / 0.22''
0.7 - 0.8 mm

2 um / 0.00008''

0.45-0.65 N

0.04 N/mm

0.7 N

0.3 um/0.000012''

0.01 um/deg. Cmm

Vacuum

-

IP40 IP40 IP40

IP50 IP50 IP50

0.1 N 0.15 N -

0.3 - 0.5 N
0.75 - 1.05 N

0.2 - 0.4 N 0.2 - 0.4 N -
0.8 - 1.8 N 0.8 - 1.8 N -

Moving mass ( without measuring
contact )

-

Protection according to IEC529

- with rubber boot

(Tolerance +/- 15%)

Measuring force

- without pressure
- low pressure
- high pressure

- upper stop
- lower stop

Limit of travel

Pneumatic lifter Pneumatic lifter Pneumatic lifter

IP64

Advertising