5 technical specifications – Bowers Group Sylvac D100S Readouts User Manual
Page 69

68
4.5 TECHNICAL SPECIFICATIONS
Accuracy using extension cables:
These measuring errors are applicable only when using D100S or D101 units without recalibration:
Normal cable
+ extension up to 5 m
: additional error
1.5 µm approx.
Normal cable
+ extension up to 10 m
: additional error
3 µm approx.
Normal cable
+ extension up to 15 m
: additional error
6 µm approx.
Normal cable
+ extension up to 25 m
: additional error
8 µm approx.
Direct cable
up to 5 m
: additional error 3 µm approx.
Direct cable
up to 10 m
: additional error 6 µm approx.
Direct cable
up to 15 m
: additional error 10 µm approx.
This is a progressive error margin and recalibration of D100S unit can considerably reduce the error.
SPECIFICATIONS
CODE NUMBERS
Construction
Type of bearing for measuring plunger
LINEAR MEASURING RANGE
Total range
Accuracy over the measuring range
Increase of measuring force
Permissible lateral force
Repetability
Zero drift
Temperature range
Cable length
Extension cables
Direct cable special length
Lifting lever
Interchangeable measuring contact
0 - 50 deg. C
1 m 50 / 59''
up to 20 m / 787''
up to 20 m / 787''
M 2.5
P10 / P10L
P25
P50
Friction bearing
3.2 g / 0.11 oz 8.7 g / 0.31 oz 14.6 g / 0.51 oz
10 mm / 0.44'' 25 mm / 1'' 50 mm / 2''
10.8 mm / 0.43'' 25.8 mm / 1.02'' 52..2 mm / 2.06''
10.4 mm / 0.41'' 25.8 mm / 1.02'' 51 mm / 2.01''
0.4 - 0.5 mm 0.8 - 0.9 mm 1.0 - 1.2 mm
1 um /0.00004'' 1.5 um / 0.00006'' 4 um / 0.00016
0.6 - 0.9 N 0.6 - 1.2 N 0.6 - 1.4 N
0.03 N/mm 0.024 N/mm 0.016 N/mm
0.6 N 0.3 N 0.25 N
0.2 um/0.000008'' 0.2 um/0.000008'' 0.4 um/0.000016''
0.02 um/deg.Cmm 0.01 um/deg.Cmm 0.01 um/deg.Cmm
Friction bearing
Friction bearing
Plunger gage
Plunger gage
Plunger gage
P5(L) / P5(L) V
Friction bearing
Plunger gage
3.2 g / 0.11 oz
5 mm / 0.2''
6.5 mm / 0.26''
5.7 mm / 0.22''
0.7 - 0.8 mm
2 um / 0.00008''
0.45-0.65 N
0.04 N/mm
0.7 N
0.3 um/0.000012''
0.01 um/deg. Cmm
Vacuum
-
IP40 IP40 IP40
IP50 IP50 IP50
0.1 N 0.15 N -
0.3 - 0.5 N
0.75 - 1.05 N
0.2 - 0.4 N 0.2 - 0.4 N -
0.8 - 1.8 N 0.8 - 1.8 N -
Moving mass ( without measuring
contact )
-
Protection according to IEC529
- with rubber boot
(Tolerance +/- 15%)
Measuring force
- without pressure
- low pressure
- high pressure
- upper stop
- lower stop
Limit of travel
Pneumatic lifter Pneumatic lifter Pneumatic lifter
IP64